ISBI 2006: IEEE 2006 International Symposium on Biomedical Imaging, April 6-9, 2006, Crystal Gateway Marriott, Arlington, Virginia, U.S.A.

Technical Program

Paper Detail

Paper:FR-AM-OS1.2
Session:PET and SPECT Image Reconstruction
Time:Friday, April 7, 09:50 - 10:10
Presentation: Oral
Title: Attenuation Correction in Cardiac SPECT Using Consistency Conditions
Authors: Ana Maria Marques da Silva; PUCRS - Instituto de Física 
 Cecil Chow Robilotta; USP - Instituto de Física 
Abstract: In this work the approach for obtaining the attenuation maps for cardiac SPECT was the estimation directly from the emission data, without transmission imaging, based on consistency conditions of the attenuated Radon transform. Several counting rates and attenuation coefficients on the corrected images were studied, using the MCAT phantom and SimSET package. Given an initial estimate of the attenuation map, the use of consistency conditions always resulted in a same form, for any values initials of the group of parameters. In spite of the fact that the error falls with the increasing counting rate, higher counts are not able to determine the best attenuation coefficient. The results shows that consistency conditions can be useful for attenuation correction, but the scattering correction is the most important factor in the quantitative reconstruction in cardiac SPECT.



ISBI 2006 is sponsored by

IEEE IEEE Signal Processing Society IEEE Engineering in Medicine and Biology Society

and organized in cooperation with

©2011 Conference Management Services, Inc. -||- email: webmaster@biomedicalimaging.org -||- Last updated Friday, February 03, 2006